Mouse Transforming Growth factor β1 (TGF-β1) ELISA Kit
Catalog No. CSB-E04726m
(96 tests)
? This immunoassay kit allows for the in vitro quantitative determination of mouse TGF-β1 concentrations in serum, plasma and other biological fluids.
? Expiration date six months from the date of manufacture
? FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC PROCEDURES.
廈門慧嘉生物經(jīng)營(yíng)ELISA試劑盒及抗體、細(xì)胞因子、生化試劑、耗材等生物試劑產(chǎn)品。誠(chéng)信經(jīng)營(yíng),價(jià)格實(shí)惠,服務(wù)周到,質(zhì)量有保證。歡迎廣告老師來(lái)詢!: :
http://www.biohj.com/download.aspx(說(shuō)明書下載)
INTRODUCTION
Transforming growth factor beta 1 or TGF-β1 is a polypeptide member of the transforming growth factor beta superfamily of cytokines. It is a secreted protein that performs many cellular functions, including the control of cell growth, cell proliferation, cell differentiation and apoptosis. TGF-β1 was first identified in human plaets as a protein with a molecular mass of 25 kilodaltons with a potential role in wound healing. It was later characterized as a large protein precursor (containing 390 amino acids) that was proteolytically processed to produce a mature peptide of 112 amino acids. TGF-β1 plays an important role in controlling the immune system, and shows different activities on different types of cell, or cells at different developmental stages. Most immune cells (or leukocytes) secrete TGF-β1. Some T cells (e.g. regulatory T cells) release TGF-β1 to inhibit the actions of other T cells. Interleukin 1-and interleukin 2-dependent proliferation of activated T cells, and the activation of quiescent helper T cells and cytotoxic T cells is prevented by the activity of TGF-β1. Similarly, TGF-β1 can inhibit the secretion and activity of many other cytokines including interferon-γ, tumor necrosis factor-alpha (TNF-α) and various interleukins. It can also decrease the expression levels of cytokine receptors, such as the IL-2 receptor to down-regulate the activity of immune cells. However, TGF-β1 can also increase the expression of certain cytokines in T cells and promote their proliferation, particularly if the cells are immature. TGF-β1 has similar effects on B cells that also vary according to the differentiation state of the cell. It inhibits proliferation and apoptosis of B cells, and plays a role in controlling the expression of antibody, transferrin and MHC class II proteins on immature and mature B cells. The effects of TGF-β1 on macrophages and monocytes are predominantly suppressive; this cytokine can inhibit the proliferation of these cells and prevent their production of reactive oxygen and nitrogen intermediates. However, as with other cell types, TGF-β1 can also have the opposite effect on cells of myeloid origin. For example, TGF-β1 acts as a chemoattractant, directing an immune response to some pathogens; macrophages and monocytes respond to low levels of TGF-β1 in a chemotactic manner. Furthermore, the expression of monocytic cytokines and phagocytic killing by macrophages can be increased by the action of TGF-β1.
PRINCIPLE OF THE ASSAY
The microtiter plate provided in this kit has been pre-coated with an antibody specific to TGF-β1. Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for TGF-β1 and Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then a TMB (3,3',5,5' tetramethyl-benzidine) substrate solution is added to each well. Only those wells that contain TGF-β1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of TGF-β1 in the samples is then determined by comparing the O.D. of the samples to the standard curve.
DETECTION RANGE
0.8 ng/ml-50 ng/ml. The standard curve concentrations used for the ELISA’s were 50 ng/ml, 25 ng/ml, 12.5 ng/ml, 6.2 ng/ml, 3.2 ng/ml, 1.6 ng/ml, 0.8 ng/ml.
SPECIFICITY
This assay recognizes recombinant and natural mouse TGF-β1. No significant cross-reactivity or interference was observed.
SENSITIVITY
The minimum detectable dose of mouse TGF-β1 is typically less than 0.2 ng/ml. The sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero.
MATERIALS PROVIDED
Reagent Quantity
Assay plate 1 Standard 2 Sample Diluent 1 x 20 ml Biotin-antibody Diluent 1 x 10 ml HRP-avidin Diluent 1 x 10 ml Biotin-antibody 1 x 120µl HRP-avidin 1 x 120µl
1 x20 ml
Wash Buffer
(25×concentrate)
TMB Substrate 1 x 10 ml
Stop Solution 1 x 10 ml
STORAGE
1 Unopened test kits should be stored at 2-8°C upon receipt and the microtiter plate should be kept in a sealed bag. The test kit may be used throughout the expiration date of the kit, provided it is stored as prescribed above. Refer to the package label for the expiration date.
2 Opened test plate should be stored at 2-8°C in the aluminum foil bag with desiccants to minimize exposure to damp air. The kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
3 A microtiter plate reader with a bandwidth of 10 nm or less and an optical density range of 0-3 OD or greater at 450nm wavelength is acceptable for use in absorbance measurement.
REAGENT PREPARATION
Bring all reagents to room temperature before use.
1 Wash Buffer If crystals have formed in the concentrate, warm up to room temperature and mix gently until the crystals have compley dissolved. Dilute 20 ml of Wash Buffer Concentrate into deionized or distilled water to prepare 500 ml of Wash Buffer.
2 Standard Centrifuge the standard vial at 6000-10000rpm for 30s. Reconstitute the Standard with 1.0 ml of Sample Diluent. This reconstitution produces a stock solution of 50 ng/ml. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions. The undiluted standard serves as the high standard (50 ng/ml). The Sample Diluent serves as the zero standard (0 ng/ml). Prepare fresh for each assay. Use within 4 hours and discard after use.
3 Biotin-antibody Centrifuge the vial before opening. Dilute to the working concentration using Biotin-antibody Diluent(1:100), respectively.
4 HRP-avidin Centrifuge the vial before opening. Dilute to the working concentration using HRP-avidin Diluent(1:100), respectively.
Precaution: The Stop Solution provided with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
OTHER SUPPLIES REQUIRED
? Microplate reader capable of measuring absorbance at 450 nm, with the correction wavelength set at 540 nm or 570 nm.
? Pipettes and pipette tips.
? Deionized or distilled water.
? Squirt bottle, manifold dispenser, or automated microplate washer.
? An incubator which can provide stable incubation conditions up to 37°C±0.5°C.
SAMPLE COLLECTION AND STORAGE
? Serum Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at 1000 g. Remove serum and assay immediay or aliquot and store samples at -20°C. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
? Plasma Collect plasma using citrate, EDTA, or heparin as an anticoagulant. Centrifuge for 15 minutes at 1000 g within 30 minutes of collection. Assay immediay or aliquot and store samples at -20°C. Centrifuge the sample again after thawing before the assay. Avoid repeated freeze-thaw cycles.
Note: Grossly hemolyzed samples are not suitable for use in this assay.
ASSAY PROCEDURE
Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate. All the reagents should be added directly to the liquid level in the well. The pipette should avoid contacting the inner wall of the well.
1 Add 100µl of Standard, Blank, or Sample per well. Cover with the adhesive strip. Incubate for 2 hours at 37°C.
2 Remove the liquid of each well, don’t wash.
3 Add 100µl of Biotin-antibody working solution to each well. Incubate for 1 hour at 37°C. Biotin-antibody working solution may appear cloudy. Warm up to room temperature and mix gently until solution appears uniform.
4 Aspirate each well and wash, repeating the process three times for a total of three washes. Wash: Fill each well with Wash Buffer (200µl) and let it stand for 2 minutes, then remove the liquid by flicking the plate over a sink. The remaining drops are removed by patting the plate on a paper towel. Complete removal of liquid at each step is essential to good performance.
5 Add 100µl of HRP-avidin working solution to each well. Cover the microtiter plate with a new adhesive strip. Incubate for 1 hour at 37°C.
6 Repeat the aspiration and wash three times as step 4.
7 Add 90µl of TMB Substrate to each well. Incubate for 10-30 minutes at 37°C. Keeping the plate away from drafts and other temperature fluctuations in the dark.
8 Add 50µl of Stop Solution to each well when the first four wells containing the highest concentration of standards develop obvious blue color. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
9 Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
CALCULATION OF RESULTS
Using the professional soft "Curve Exert 1.3" to make a standard curve is recommended, which can be downloaded from our web.
Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the TGF-β1 concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
LIMITATIONS OF THE PROCEDURE
? The kit should not be used beyond the expiration date on the kit label.
? Do not mix or substitute reagents with those from other lots or sources.
? It is important that the Standard Diluent selected for the standard curve be consistent with the samples being assayed.
? If samples generate values higher than the highest standard, dilute the samples with the appropriate Standard Diluent and repeat the assay.
? Any variation in Standard Diluent, operator, pipetting technique, washing technique, incubation time or temperature, and kit age can cause variation in binding.
? This assay is designed to eliminate interference by soluble receptors, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Quantikine Immunoassay, the possibility of interference cannot be excluded.
TECHNICAL HINTS
? Centrifuge vials before opening to collect contents.
? When mixing or reconstituting protein solutions, always avoid foaming.
? To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
? When using an automated plate washer, adding a 30 second soak period following the addition of wash buffer, and/or rotating the plate 180 degrees between wash steps may improve assay precision.
? To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary.
? Substrate Solution should remain colorless or light blue until added to the plate. Keep Substrate Solution protected from light. Substrate Solution should change from colorless or light blue to gradations of blue.
? Stop Solution should be added to the plate in the same order as the Substrate Solution. The color developed in the wells will turn from blue to yellow upon addition of the Stop Solution. Wells that are green in color indicate that the Stop Solution has not mixed thoroughly with the Substrate Solution.