Voltage Calibrator to 1000 V Extremely User Friendly
Power Calibrator to 2400 W NIST Traceable
參考價 | 面議 |
更新時間:2017-11-21 11:22:55瀏覽次數(shù):610
聯(lián)系我們時請說明是化工儀器網(wǎng)上看到的信息,謝謝!
Specifications
DC voltage
range | % of value + % of range |
0 µV - 20 mV | 0.05 + 0.0 + 10 µV |
20 mV - 200 mV | 0.01 + 0.0 + 10 µV |
200 mV - 2 V | 0.003 + 0.0008 |
2 V - 20 V | 0.003 + 0.0005 |
20 V - 240 V | 0.003 + 0.0005 |
240 V - 1000 V | 0.005 + 0.005 |
AC voltage
range | % of value + % of range | % of value + % of range | % of value + % of range |
20 Hz - 10 kHz | 10 kHz - 50 kHz | 50 kHz - 100 kHz | |
1 mV - 20 mV | 0.2 + 0.05 + 20 µV | 0.20 + 0.10 + 20 µV | 1.0 + 0.10 + 20 µV |
20mV - 200mV | 0.1 + 0.03 + 20 µV | 0.15 + 0.05 + 20 µV | 0.3 + 0.05 + 20 µV |
200 mV - 2 V | 0.025 + 0.005 | 0.05 + 0.01 | 0.2 + 0.05 |
2 V - 20 V | 0.025 + 0.005 | 0.05 + 0.03 | 0.2 + 0.05 |
20 V - 240 V | 0.025 + 0.010 | - - | - - |
240 V - 1000 V | 0.03 + 0.02 * | - - | - - |
* valid for f < 1000 Hz |
DC current
range | % of value + % of range |
1 µA - 200 µA | 0.05 + 0.0 + 20 nA |
200 µA - 2 mA | 0.02 + 0.005 |
2 mA - 20 mA | 0.01 + 0.003 |
20 mA - 200 mA | 0.01 + 0.003 |
200 mA - 2 A | 0.015 + 0.005 |
2 A - 20 A | 0.02 + 0.010 |
AC current
range | % of value + % of range | % of value + % of range |
20 Hz - 1 kHz | 1 kHz - 5 kHz | |
1 µA - 200 µA | 0.15 + 0.0 + 20 nA | 0.30 + 0.10 + 20 nA |
200 µA - 2 mA | 0.07 + 0.01 | 0.20 + 0.05 |
2 mA - 20 mA | 0.05 + 0.005 | 0.20 + 0.05 |
20 mA - 200 mA | 0.05 + 0.005 | 0.20 + 0.05 |
200 mA - 2 A | 0.05 + 0.005 | - - |
2 A - 20 A | 0.10 + 0.03 | - - |
When option 130-50 Current coil is used, add accuracy 0.3 % of the set current to the value specified in the above table. Output current is multiplied by factor 50. |
Resistance (4-wire)
range | % of value |
0 Ohm - 100 Ohm | 0.03 + 10 mOhm |
100 Ohm - 400 Ohm | 0.015 |
400 Ohm - 2 kOhm | 0.015 |
2 kOhm - 10 kOhm | 0.015 |
10 kOhm - 40 kOhm | 0.015 |
40 kOhm - 200 kOhm | 0.015 |
200 kOhm - 1 MOhm | 0.05 |
1 MOhm - 4 MOhm | 0.1 |
4 MOhm - 20 MOhm | 0.2 |
20 MOhm - 50 MOhm | 0.5 |
Maximum allowed voltage on the load is 8 Vpk. |
Capacitance (4-wire)
range | % of value |
900 pF - 2.5 nF | 0.5 + 15 pF |
2.5 nF - 10 nF | 0.5 + 5 pF |
10 nF - 50 nF | 0.5 |
50 nF - 250 nF | 0.5 |
250 nF - 1 µF | 0.5 |
1 µF - 3.5 µF | 1.0 |
3.5 µF - 5 µF | 1.0 |
5 µF - 10 µF | 1.5 |
10 µF - 50 µF | 2.0 |
Maximum allowed voltage on the load is 8 Vpk. |
Frequency
type | PWM (pos, neg, sym) | HF * |
range | 0.1 Hz - 100 kHz | 0.1 Hz - 20 MHz |
frequency unc. (%) | 0.005 | 0.005 |
amplitude | 1 mV - 10 V | 2 V |
amplitude unc. (%) | 0.1 | 10 |
ratio | 0.00 - 1.00 | - - |
ratio unc.(%) | 0.05 | - - |
* Rise time < 5 ns |
DC power, energy
quantity | range | % of value + % of range |
voltage | 200 mV - 240 V | see voltage table |
current | 2 mA - 10 A | 0.05 + 0.01 |
AC power, energy
quantity | range | accuracy |
voltage | 200 mV - 240 V | see voltage table |
current | 2 mA - 10 A | 0.05 % + 0.01 % |
frequency | 40 Hz - 400 Hz | 0.005 |
power factor | -1 - +1 | 0.005 - 0.0005 |
phase | 0 - 360 ° | 0.15 ° - 0.25 ° |
Setting time in energy mode 10 s - 1999 s. |
Resistance temperature sensor simulation
types | Pt 1.385, Pt 1.392, Ni |
range of R0 | 20 Ohm - 2 kOhm |
range of temperature | -200 °C - 850 °C |
temperature accuracy | 0.04 °C - 0.5 °C |
temperature scale | ITS 90, PTS 68 |
Thermocouple simulation
types | R, S, B, J, T, E, K, N |
range of temperature | -250 °C - 1820 °C |
temperature accuracy | 0.4 °C - 4.0 °C |
temperature scale | ITS 90, PTS 68 |
Multimeter
quantity | range | accuracy |
VDC (DC voltage) | 0 - ± 12 V | 0.01 % + 100 µV |
mVDC (DC voltage) | 0 - ± 2000 mV | 0.01 % + 10 µV |
mADC (DC current) | 0 - ± 25 mA | 0.02 % + 1 µA |
FREQ (Frequency) | 1 Hz - 15 kHz | 0.005 % |
R4W (Resistance) | 0 - 2 kOhm | 0.02 % + 100 mOhm |
TRTD (RTD sensors) | -150 - +600 °C | 0.1 °C |
TTC (TC sensors) | -250 - +1820 °C | 0.4 - 4 °C |
SGS (strain gauge sensor) * | depends on sensor | 0.01 % + 10 µV + sensor unc. |
* Supplying voltage 2 to 10 V DC, max. current 40 mA, input resistance > 100 MOhm |
Accessories included
Power line cable | 1 pc | |
Operation manual | 1 pc | |
Option 10/11 Test cable | 2 pcs | |
Option 40 Test cable for built-in multimeter, 2 banana terminals for DC voltage and current measurement. | 1 pc | |
Option 60 Test cable for built-in multimeter, 4 banana terminals for resistance measurement as well as Pt/Ni sensor temperature measurement. | 1 pc | |
Option 70 4-wire resistance adapter. The adapter provides a four wire return to the calibrator for making resistance measurements. | 1 pc |
Optional Accessories
Option 130-50 | Current coil 50 turns. Suitable for clamp-on Current transformer testing. | |
Option 140-01 | Cable adapter for multimeters. The adapter provides support for the meter under test as well as a simple cable connection to the calibrator front panel terminals. The Adapter is equipped with a built-in Pt-100 temperature sensor. The internal meter can measure ambient temperature and display it on the screen and automatically make cold junction compensation when the TC temperature sensor simulation is selected. | |
Option 140-41 | Cable adapter for simultaneous applications. Cable adapter is designed for low-voltage/current aplications to 20 VDC and 200 mADC. There are two fields on the adapter. The first contains terminals with the calibrator output signals, the second contains internal multimeter input terminals. Various calibration tasks can be fulfilled with the adapter, especially measurements where simultaneous generation of the calibrated signal and response of unit under test are required. With the adapter, temperatures of external TC temperature sensors can be measured and displayed. | |
Option 140-02 | Set of cables and adapters. It includes Option 140-01, Option 140-41, Option 10 2pcs, Option 11 2pcs, Option 20, Option 30, Option 40, Option 60, Option 70 | |
Option 10 | Test cable 20A/1000V (black) | |
Option 11 | Test cable 20A/1000V (red) | |
Option 20 | Test cable BNC/BNC 1m | |
Option 30 | Test cable BNC/banana 1m | |
Cable GPIB | IEEE-488, 2m | |
Cable RS-232 | RS-232, 2m |
Description
The Multifunction Calibrator Model 8080 functions as a Voltage Calibrator, Current Calibrator, Power Calibrator, Resistance Calibrator, Capacitance Calibrator and a Thermocouple Calibrator. Having very low uncertainties, the Multifunction Calibrator is ideal as a standard of these electric quantities in calibration laboratories. The Multifunction Calibrator also has the feature of supplying several different waveshapes in additional to the traditional sinusoidal output. This feature makes it ideal for testing meters which are true rms reading or crest factor sensitive. These signals are also suitable for the basic calibration of oscilloscopes. The Multifunction Calibrator has the capability of measuring temperature with TC and RTD temperature sensors. A built-in multimeter can be used independently or in simultaneous operation with the Multifunction Calibrator output.
Basic Functions
In the Voltage Calibrator mode, the Multifunction Calibrator generates DC/AC voltages up to 1000 V and in the Current Calibrator mode it generates DC/AC currents up to 20 A . Using a 50-turn transformer, the AC current range can be extended to 500 A. The best accuracy of the calibrator on DC voltage ranges is 0.0035%, on AC voltage ranges 0.03%, on DC current ranges 0.013% and on AC current ranges 0.055%. The frequency range is from 20 Hz to 100 kHz. The calibrator can generate periodic non-harmonic signals with a defined crest factor. Such signals can be used to test the sensitivity of multimeters to distorted signals. The Multifunction Calibrator simulates both resistance and capacitance. The resistance range extends to 50 MOhm and the capacitance range is from 1 nF to 50 µF. The precision of both quantities is sufficient for the calibration of commonly used hand-held multimeters. The basic accuracy of the resistance ranges and capacitance ranges is 0.03%, and 0.5% respectively. The resistance calibration may be used with both DC and AC signals up to 1 kHz.
The frequency function of the calibrator makes it possible to generate a square wave signal with adjustable and calibrated duty cycle, frequency and amplitude in the range from 1 mV to 10 V in the frequency band up to 10 kHz. In the HF mode a square wave signal up to 20 MHz with a very low risetime can be generated. The frequency function is suitable for the calibrations of corresponding frequency ranges of multimeters, and for calibration of the channel sensitivities and time bases of the oscilloscopes.
In the Power Calibration mode, the Model 8080 can be used as a source to calibrate DC and single-phase AC watt meters, power analyzers, and energy meters. In this mode, the output voltage can be set up to 240 V, the output current up to 10A, and the the power factor in range from -1 to +1 in the frequency band from 20 Hz to 400 Hz. The current capability of the voltage output is 30 mA so that analog watt meters, which usually have lower input impedances at their voltage input, may be calibrated.. For calibration of thermometers and temperature regulators, the temperature function is simulated by the Multifunction Calibrator. It is able to simulate all commonly used Pt and Ni resistance sensors as well as TC sensors of the R, S, B, J, T, E, K and N types. Compensation of the TC cold junction is made either by entering a value from the keyboard, or automatically by measuring the ambient temperature with a Pt-100 sensor. The precision of the simulated resistance and TC sensors depends on set value and type of the sensor. For resistance sensors the accuracy band is in the range from 0.04 °C to 0.5 °C, for TC sensors from 0.4°C to 4.0 °C.
Built-in multimeter
An internal multimeter with basic ranges of 20 mA, 200 mV, 10 V and 2 kOhm is a standard feature off the Multifunction Calibrator. With an accuracy of 0.01%, it enables the user to measure the output signals of various types of transducers. With external strain gauge sensors, pressure, torsion, strength, etc. can be measured and displayed.
Calibrator - Tester
The Multifunction Calibrator can be used both as a precision signal source for a device under test and as a meter to measure the resultant voltages generated by the device. Programmable capability of the calibrator enables the user to set 10 steps of a testing procedure. The testing can run automatically, and after it is completed, PASS/FAIL status is displayed. With the testing function, isolated relay contacts activated by the FAIL indicator, make it possible to control other equipment.
User friendly
The Model 8080 Calibrator is equipped with a number of other functions which make it extremely easy to use. Among them, are the capability to set relative deviations from the actual value of the selected output signal, the continuous displaying of the output signal accuracy, and the internal calibration procedure. The large, easy to read, comprehensive luminescent display contains all of the necessary information to set up any of the Multifunction Calibrator's functions. In addition, frequently used functions have specific keys assigned to them. The calibrator is equipped with the IEEE-488.2 interface and with the RS-232 serial interface.
Operation
The operation of the Model 8080 display, front panel keyboard and front panel terminals can be seen iillustrated by clicking on the links below.