詳細介紹
康晨希供應MESALABS BD測試包
♦ Air Removal Test Pack – Intended to evaluate the performance of the Air Removal (vacuum) system of Pre-Vacuum equipped Steam Sterilizers.
空氣排出測試包-評估預真空蒸汽滅菌器的空氣排除(真空)性能。
♦ Pre-assembled, disposable test pack. After exposure, unwrap pack and remove the indicator sheet for your records.
預裝配的一次性測試包。滅菌后,解開包裹,取出試紙記錄結(jié)果。
♦ 510(k) notification and meets AAMI, EN-867-3 and ISO 11140 air removal test.
符合510(k),AAMI,EN-867-3和ISO11140空氣排除測試標準。
♦ Lead Free Ink on Indicator Sheet – Safer for medical personnel and the environment.
試紙采用無鉛油墨-保證醫(yī)療人員和環(huán)境的安全。
♦ Distinct, non-reversible pink to black color change – easy to interpret.
*的不可逆的粉紅色到黑色的變化-容易判讀。
滅菌前 滅菌成功 滅菌失敗
Procedures for Use
- Test sterilizer once a day.
- Test at the beginning of each day. A shortened cycle should be used to preheat the system (i.e. omit the drying phase). If the sterilizer is in continual use, then testing can be performed at any time, but it should be tested at the same time every day.
- Place the pack on the bottom rack in the front section of the sterilizer, near the door and over the drain.
- The test is most accurate when nothing else is in the chamber, so do not use this cycle to sterilize items or packs. Other items in this cycle may invalidate the test.
- Run a pre-vacuum cycle following the parameters printed on the indicator instructions. You may eliminate the drying phase to save time without affecting the test.
- After the cycle is complete, remove the pack and open to examine the indicator sheet. CAUTION! Test pack may still be hot so use caution when handling to avoid injury.
- Record the necessary information on the sheet (or in a log book) and store for your records.
- The filter paper pack can be discarded into a paper recycle bin.
產(chǎn)品實物圖:康晨希供應MESALABS BD測試包