請輸入產(chǎn)品關(guān)鍵字:
郵編:100020
聯(lián)系人:周經(jīng)理
電話:010-58790591
傳真:010-58790595
手機:18510328756
留言:發(fā)送留言
網(wǎng)址:www.upwardstek.cn
商鋪:http://m.facexiu.com/st3671/
NIBP100D-BIOPAC NIBP100D *血壓 采集記錄系統(tǒng)
【產(chǎn)品簡介】
特點:
★-*、操作簡便、連續(xù)的實時數(shù)據(jù)、波形顯示;
★-波形與數(shù)據(jù)可本機打印輸出;
★-也可以接入生理記錄儀通過PC和顯示器做到實時監(jiān)測、波形數(shù)據(jù)保存、事后分析統(tǒng)計、Excel★-表格成檔導(dǎo)出。
【詳細(xì)說明】
Continuous non-invasive arterial pressure shows high
accuracy in comparison to invasive intra-arterial blood
pressure measurement
I. INTRODUCTION
Continuous blood pressure (BP) monitoring is required in
a multitude of clinical settings, especially in perioperative
care. For inpatient surgeries, the American Society
of Anesthesiologists (ASA) requires continuous perioperative
blood pressure monitoring at least for patients with
severe systemic disease; this necessitates the invasive
placement of an intra-arterial catheter. In all other cases
intermittent non-invasive blood pressure monitoring (NBP)
is the standard of care. Therefore, the patients` blood
pressure may not be monitored at all times.
A recent representative survey(1) among Austrian and
German physicians (N=198) provides evidence that, in
82% of inpatient surgeries, non-invasive blood pressure
monitoring is used. However, in 25% of these cases,
especially in surgeries where hemodynamic instabilities
can be expected or where aggressive management of
blood pressure might be required (e.g. in urologic, extended
laparoscopic, orthopaedic or vascular surgeries, in
surgeries in gynecology and obstetrics, in medium to extended
intestinal surgery and elective or urgent trauma
surgery(2)), anesthetists would prefer a non-invasive continuous
blood pressure monitoring to have better control
over the patient’s hemodynamics. In the remaining 18%
of inpatient surgeries, BP is measured continuously using
invasive catheters (IBP), mainly in patients where cardiovascular
instability is expected and thus ASA guidelines
specifically require continuous BP measurement and/or
where repeated blood gas analysis is needed. Note that,
in 26% of these cases the invasive catheter is inserted
only to enable continuous blood pressure monitoring.
However, this is a time-consuming and cost-intensive
procedure, causing pain for the patient and including
the risk of infection, and thus should be replaced by a
non-invasive method if possible.
There are a number of studies stressing the importance
of continuous perioperative blood pressure monitoring:
e.g., more than 20% of all hypotensive episodes during
surgeries may be missed by intermittent upper-arm
blood pressure readings and another 20% may be detected
with a delay(3). This in turn may prevent immediate
treatment or even lead to missing complete hypotensive
episodes. It has been shown that intraoperative
hypotension preceeds 56% of perioperative cardiac
arrests(4) and is associated with a significant increase of
the 1-year post surgical mortality rate(5), indicating that
intermittent NBP monitoring can be insufficient.
Consequently, there seems to be a discrepancy between
the number of cases where continuous blood
pressure monitoring is needed and those cases where
it is actually used: Due to its invasive nature and associated
risks, intra-arterial catheters can only be justified in
a limited number of patients whereas anesthetists would
like to perform risk-free continuous BP monitoring in a greater
number of cases. For exactly these situations CNAP™
has recently become available(6, 7, 8, 9, 10, 11). CNAP™ is designed
for anesthetists who look for more control in situations
when continuous blood pressure is desirable, but
the risks and burden of an arterial line are not justified.
CNAP™ provides continuous, non-invasive and risk-free
beat-to-beat blood pressure measurement.
The aim of the present report was to evaluate the accuracy
of CNAP™ in a real-life perioperative setting by
comparing simultaneous measurements of CNAP™ to
continuous intra-arterial pressure monitoring.
II. Methods
Data recording
The measurements were conducted in a perioperative
setting at the Department of Anesthesiology at Landeskrankenhaus
Bruck an der Mur (Austria). In all patients
included in this report, continuous BP monitoring was
indicated by clinical safety standards. Arterial pressure
was measured simultaneously with an invasive catheter
(Edwards Life Sciences™ Pressure Monitoring Set, Irvine,
USA, connected to Datex Ohmeda S/5 monitor, GE, Helsinki,
Finland) and the CNAP™ Monitor 500i (CNSystems
Medizintechnik AG, Graz, Austria) in fifteen patients undergoing
orthopedic, cardiac and vascular surgeries
(seven female and eight male patients, mean age of
71 years, range 33 to 82 years, ASA classifications I-III: I in
1 case, II in 12 cases, III in 2 cases). The arterial catheter
was placed ipsi-laterally (n=5) or contra-laterally (n=10)
to the CNAP™ finger cuff in the A. radialis or A. brachialis,
depending on indication and requirements. The surgery
durations averaged 1h39min with a minimum of 44min
and a maximum of 3h01min, the total duration of recordings
obtained was approx. 25 hours.
Data processing
From the IBP as well as from the CNAP™ signal, systolic,
diastolic and mean pressure values were derived for
each second. If one of the signals was missing (e.g. due
to transmission faults or artifacts) for one data point, all
other measurements for that data point were consequently
discarded. Otherwise, no further data processing
was performed and a total of 75,485 data points
were included into the statistical comparison.
Sackl-Pietsch E., Department of Anesthesiology, Landeskrankenhaus Bruck an der Mur, Austria
Data comparison
For a comprehensive evaluation of CNAP™, its underlying
mechanisms have to be considered: CNAP™ is an
integrated solution where relative BP changes are measured
at the finger sensor which are turned into absolute
values based on initial readings from its integrated
NBP-unit. This fact needs to be taken into consideration
when comparing the blood pressure readings recorded
by CNAP™ and IBP.
Since three measurement positions are combined in this
comparison (CNAP™ finger sensor, CNAP™ NBP-unit and
IBP catheter), some physiological facts have to be taken
into account: namely, transformations of BP amplitudes
and waveforms as illustrated in figure 1. This implies that
a systematic offset between CNAP™ and IBP can be expected.
Thus, it is not surprising that even the AAMI-SP10 standard
recommended by the FDA reports substantial differences
between indirect NBP and direct intra-arterial
measurements(12). A meta-analysis with the results of nine
studies totaling 330 patients was performed which quantifies
this systematic offset: The average differences between
arterial and NBP-cuff systolic BP ranged from 0.8 to
13.4 mmHg with standard deviations (SD) ranging from 0
to 13.0 mmHg. Diastolic BP showed average differences
from 0.8 to 18.0 mmHg with SDs ranging from 0.0 to 10.2
mmHg.
This offset may be even magnified when IBP and NBP
recordings are taken on contra-lateral arms. Note that in
10 out of the 15 patients reported on here, CNAP™ and
IBP were placed on contra-lateral arms.
Therefore, the following differences between CNAP™
and IBP can be expected:
(i) Differences between the two BP waveforms.
(ii) The characteristic offset between the absolute values
of systolic, diastolic and mean pressure.
Figure 1: Different blood pressure waveforms and amplitudes in
the (1) A. brachialis, (2) A. radialis and (3) A. digitalis, resulting
in different systolic and diastolic values
III. RESULTS
Waveform comparison
Figure 2 shows blood pressure waveforms recorded by
CNAP™ compared directly to intra-arterial blood pressure
waveforms. The upper graph shows a short episode
of stable blood pressure. The bottom-up arrows indicate
rising and the top-down arrows indicate falling BP ramps
considered as results of volume status, the Frank-Starling
mechanism and autonomic regulation(13). The lower graph
shows BP changes caused by perioperative treatment
or patient movement. Due to the fact that data
was recorded in the clinical routine, no further information
about the patient`s treatment at this special time
slice is available. Nevertheless, the good accordance of
waveforms indicates that CNAP™ can follow fast blood
pressure variations changes as well as IBP.
Hemodynamic changes
For clinical application it is important to ensure that
CNAP™ is able to monitor fast hemodynamic changes.
In figure 3 an example is displayed where short-term hemodynamic
variability during 25 minutes of orthopedic
surgery can be observed clearly: CNAP™ and IBP display
a parallel hemodynamic trend with the typical offset
between indirect and direct measurement methods.
Sackl-Pietsch E., Department of Anesthesiology, Landeskrankenhaus Bruck an der Mur, Austria 2
Continuous non-invasive arterial pressure shows high accuracy in comparison to invasive intra-arterial
blood pressure measurement
Figure 2: Blood pressure tracings showing the agreement of
CNAP™ (solid line) with IBP (dotted line) during anesthesia.
Boxplots for all patients’ data sets
Figure 4 shows boxplots for all 15 data sets, for mean
BP values. This graph illustrates that most of the patients
show a characteristic offset between CNAP™ and IBP.
Bland-Altman-plots for the complete data set
The differences of CNAP™ and IBP data points were
computed for every data point (n = 75,485) and plotted
vs. their average, resulting in the Bland-Altman-plot of
Figure 5. No distinct trend of blood pressure difference
in relation to the absolute mean values of pressure can
be detected, i.e. the diffe difference between the two
recording methods is the same over the whole range of
values.
Furthermore, table 1 shows mean values and standard
deviations of differences of CNAP™ to IBP for systolic,
mean and diastolic pressure for each patient separay
as well as for the whole sample.
Figure 3: Comparison of short-term trends of systolic, diastolic
and mean blood pressure measurements from CNAP™ (solid
lines) and from IBP (dotted lines) during 25 min of anesthesia.
Figure 4: Boxplots of differences between CNAP™ and IBP
values for all 15 patients (mean BP [mmHg]). The boxes contain
the middle 50% of the data, the horizontal lines show the
median. The upper and lower edges of the boxes indicate the
75th and 25th percentiles, respectively. The 5-95% range of the
data is indicated by the ends of the vertical lines.
Figure 5: Bland-Altman-plot of differences vs. average of all
data points (CNAP™ vs. IBP values, n=75,485) for mean BP
[mmHg].
Systolic BP Mean BP Diastolic BP
patient mean SD mean SD mean SD
1 -10,03 13,83 4,29 9,87 8,80 6,80
2 2,56 7,54 16,09 5,82 19,24 5,88
3 -2,81 7,17 6,99 6,51 12,27 7,21
4 -7,82 12,06 1,88 12,62 9,94 14,11
5 1,31 6,63 14,41 5,88 20,25 4,70
6 -16,43 5,11 -9,44 4,15 -3,99 4,38
7 -1,33 8,00 5,44 6,15 14,46 5,07
8 -10,77 5,69 1,91 3,71 7,34 2,86
9 -11,20 7,78 -0,81 6,71 3,75 6,91
10 -9,93 7,82 1,93 3,86 7,16 3,22
11 -25,82 8,37 -7,48 4,62 0,22 3,89
12 -1,45 6,95 6,52 7,73 10,95 6,41
13 0,24 11,62 6,81 7,91 11,09 7,34
14 33,55 4,59 32,00 7,10 37,77 5,77
15 2,89 10,49 13,58 5,84 19,69 4,99
Total -2,96 13,81 6,66 11,23 12,36 10,91
TABLE 1: Means and standard deviations (SD) of differences
between CNAPTM and IBP [mmHg].
Sackl-Pietsch E., Department of Anesthesiology, Landeskrankenhaus Bruck an der Mur, Austria 3
Continuous non-invasive arterial pressure shows high accuracy in comparison to invasive intra-arterial
blood pressure measurement
IV. DISCUSSION
Within an every day clinical setting, CNAP™ and IBP readings
were recorded simultaneously during inpatient surgeries.
The results of this perioperative comparison indicate
that CNAP™ has a high usability during anesthetic
care: the overall statistical analyses of systolic, mean and
diastolic blood pressure show small differences and standard
deviations between the two methods. The graphical
comparison of BP waveforms and short-term trends
during anesthesia indicates that CNAP™ can follow hemodynamic
variability as fast as IBP. These results give
strong support to a high accuracy of the non-invasive
CNAP™ device in comparison to the invasive measurement.
The waveforms of CNAP™ and IBP shown in figure 2 comply
well with the physiological expectations (see section “Methods”).
As can be seen, CNAP™ corresponds to the IBP
signal both in resting conditions as well as in movement.
For perioperative usability of the CNAP™ system, it is essential
to show that CNAP™ can deal with hemodynamic
changes as well as IBP: The trends of systolic, diastolic
and mean BP depicted in figure 3 show excellent visual
accordance between the two devices.
To illustrate the overall agreement between CNAP™ and
IBP, figures 4 and 5 sum up the results for all 15 patients.
The validation of CNAP™ with a total observation duration
of about 25 hours and 75,485 data points is very acceptable:
The mean values and standard deviations of
differences to the intra-arterial recordings comply with
the results of the meta-analysis recommended by the
FDA.
As can be seen in figure 4, all patients have their own
characteristic offset between CNAP™ and IBP. Only patient
no. 14 seems to slightly deviate from the rest with a
higher pressure difference which may be explained by
the patients’ arteries: in patient no. 14 the vessels were
described by the clinician as ‘stiff’ and the IBP readings
as ‘dependent on bedding’, thus making the arterial reference
less reliable and the results surprisingly good. On
the other hand, not even in the case where a patient’s
peripheral perfusion was described by the physician
as “poor” (patient no. 11) did the CNAP™ system fail to
quickly find a suitable BP waveform and the results compared
to IBP are very satisfactory.
The individual, physiologically-determined offset can
also be seen clearly in the cluster of data points of each
patient in figure 5 (e.g., note patient no. 14 in the upper
right-hand corner). Nevertheless, the Bland-Altman-plot
between CNAP™ and IBP shows no distinct trend of mean
pressure difference in relation to the average values of
pressure, i.e. the difference between the two recording
methods is the same over the whole range of values. This
indicates that CNAP™ measurement is reliable in normal,
hypotensive and hypertensive episodes.
The mean values and standard deviations of differences
between CNAP™ and IBP reported in table 1 confirm the
findings of the meta-analysis in the current ANSI standard.
These results are very satisfactory considering the patient
sample included in this report. Note that data was recorded
in patients with severe systemic disease or during
higher-risk surgeries where the placement of an invasive
catheter was motivated by safety considerations.
Although the results of this report indicate a high clinical
usability of CNAP™, some remarks have to be made
about the comparison to IBP measurements: There is
common agreement that “true” blood pressure is best
determined directly using a reliable, calibrated transducer
in an artery. Nevertheless, there is also consensus
that the direct intra-arterial measurement is fraught with
its inherent set of issues, including variability with radial
position, vasoconstriction, the effects of flow-velocity
changes and the frequency response of amplifier and
transducer. Taking this into account, the results of this
present report are even more remarkable.
V. Conclusion
On the whole, the reported results provide clear evidence
of an excellent clinical feasibility and high accuracy
of the non-invasive BP measurement device CNAP™
in comparison to IBP.
With intermittent measurement of oscillometric sphygmomanometers
(NBP), short-term but clinically relevant
hemodynamic changes during anesthesia are not satisfactorily
detectable. Therefore, the demand from
anesthetists for a system providing non-invasive, continuous
beat-to-beat BP is increasing.
CNAP™ provides patient comfort and usability similar to
a standard upper-arm NBP and clinical data shows that
its accuracy is comparable to IBP. Thus, CNAP™ is the
convenient solution for anesthetists who want to have
comprehensive hemodynamic control to ensure highest
patient safety.
Sackl-Pietsch E., Department of Anesthesiology, Landeskrankenhaus Bruck an der Mur, Austria 4
Continuous non-invasive arterial pressure shows high accuracy in comparison to invasive intra-arterial
blood pressure measurement
VI. REFERENCES
1. von Skerst B: Market survey, N=198 physicians in Germany
and Austria, Dec.2007 - Mar 2008, InnoTech Consult
GmbH, Germany
2. Ezekiel MR. Handbook of Anesthesiology. Current Clinical
Strategies Publishing. 2003
3. Dueck R, Jameson LC. Reliability of hypotension detection
with noninvasive radial artery beat-to-beat versus upper
arm cuff BP monitoring. Anesth Analg 2006, 102 Suppl: S10
4. Sprung J, Warner ME, Contreras ME et al. Predictors of
Survival following Cardiac Arrest in Patients Undergoing
Noncardiac Surgery. Anesthesiology 2003; 99:259–69
5. Monk TG, Saini V, Weldon BC, Sigl JC. Anesthetic management
and one-year mortality after noncardiac surgery.
Anesth Analg. 2005 Jan;100(1):4-10.
6. Fortin J, Gratze G, Wach P, Skrabal: Automated noninvasive
assessment of cardiovascular function, spectra
analysis and baroreceptor sensitivity for the diagnosis
of syncopes. World Congress on Medical Physics and
Biomedical Engineering. Med & Biol Eng & Comput, 35,
Supplement I, 466 (1997).
7. Gratze G, Fortin J, Holler A, Grasenick K, Pfurtscheller G,
Wach P, Kotanko P, Skrabal F: A software package for
non-invasive, real time beat to beat monitoring of stroke
volume, blood pressure, total peripheral resistance and for
assessment of autonomic function. Comp in Bio & Medicine;
28, 121-142 (1998).
8. Fortin J, Habenbacher W, Gruellenberger R, Wach P, Skrabal
F: Real-time Monitor for hemodynamic beat-to-beat
parameters and power spectra analysis of the biosignals.
Proc. of the 20th Annual International Conference of the
IEEE Eng in Medicine and Biology Society, 20, 1 (1998).
9. Fortin J, Marte W, Grüllenberger R, Hacker A, Habenbacher
W, Heller A, Wagner Ch, Wach P, Skrabal F: Continuous
non-invasive blood pressure monitoring using concentrically
interlocking control loops. Computers in Biology
and Medicine 36 (2006) 941–957
10. Fortin J, Alkan S, Wrede C E, Sackl-Pietsch E and Wach P:
Continuous Non-invasive Arterial Pressure (CNAP™) – An
Innovative Approach of the Vascular Unloading Technique.
Submitted for publication in Blood Pressure April
2008.
11. Fortin J: Continuous Non-invasive Measurements of Cardiovascular
Function. PhD-thesis, Institute of Biomedical Engineering,
University of Technology Graz, 2007, pp. 103-21.
12. Association for the Advancement of Medical Instrumentation.
American National Standard. Manual, electronic or
automated sphygmomanometers ANSI/AAMI SP10-2002/
A1. 3330 Washington Boulevard, Suite 400, Arlington, VA
22201-4598, USA: AAMI; 2003
13. Parati G, Omboni St, Frattola A, Di Rienzo M, Zanchetti A,
Mancia G. Dynamic evaluation of the baroreflex in ambulant
subject. In: Blood pressure and heart rate variability,
edited by di Rienzo et al. IOS Press, 1992, pp. 123-137.
Sackl-Pietsch E., Department of Anesthesiology, Landeskrankenhaus Bruck an der Mur, Austria 5