Blood Pressure performance testing device,
BPA700
NIBP Simulator & Analyzer is a performance testing device that can be used standalone or connected to a PC. It contains wide display range, arbitrarily adjustable pulse envelope to set precise values. Compatible with all types of blood pressure monitors, benefits R&D and compliance testing
Compliant test for IEC 80601-2-30, YY9706.230 and comply with simulator standard ISO/TS 81060-5
SPECIFICATION
ParametersSpecifications
Pressure UnitmmHg, kPa
ManometerRange: 0 to 400mmHg
Resolution: 0.1mmHg
Accuracy: ± (0.3% of reading + 0.5mmHg)
Pressure Source Target Pressure: 20 to 400mmHg
Settle time: 5seconds
Resolution: 1mmHg
Accuracy: ± 0.5mmHg
Pulse Rate
Range: 30 to 300bpm
Accuracy: ± 1bpm
Pulse Amplitude: 2mmHg max
(500ml air réservoir)
Pulse Volume: 0 to 2ml
WaveformOscillometric
Leak Test Leak Test Time: 0 to 300seconds
Target Pressure: 20 to 400mmHg
Range: 0 to 300mmHg/min
Overpressure Test Release Time: 0 to 300seconds
Range: Automatic inflation
Instantaneous: 0 to 400mmHg
BP Envelope Shift Systolic Range: ± 20mmHg
Diastolic Range: ± 20mmHg
BP Dynamic Range Diastolic Range: 10 to 250mmHg
Systolic Range: 25 to 300mmHg
Dynamic NIBP Simulation Repeatability ±2mmHg
0.05mmHg (Standard deviation)
Self-Leakage<1mmHg/min
(500ml air réservoir volume)
Environment Operating Temperature: 10℃ to 40℃
Storage Temperature: 0℃ to 50℃
Humidity: 0–90% RH (Non-condensing)
Shape Dimensions: 326.4 x 315 x 88mm
Display: Graphic LCD Resolution 320 x 240pixels
Weight: 4.5kg
CommunicationUSB
PowerAC 100-240V,50/60Hz
Blood Pressure performance testing device